Safe use of tacrolimus relies on regular whole-blood drug monitoring. Of the methods used to assess whole-blood tacrolimus concentration, antibody-conjugated magnetic immunoassay is mostly used for therapeutic drug monitoring because it requires only a minimal sample preparation and no pretreatment procedure. However, several cases recently have been reported in which abnormally false elevated tacrolimus concentrations were measured by antibody-conjugated magnetic immunoassay (>15 ng/mL), despite the absence of clinical symptoms.
We present 2 cases of falsely detected tacrolimus concentrations that did not show abnormally high values within the therapeutic range. Whole-blood tacrolimus concentrations obtained by antibody-conjugated magnetic immunoassay showed well-controlled concentrations (approximately 2-8 ng/mL), whereas those obtained by another immunoassay and in washed erythrocytes were below the assay range (< 1.2 ng/mL). Thus, antibody-conjugated magnetic immunoassay can elicit falsely positive results of tacrolimus concentrations, even though they are within the therapeutic range.
Key words : Liver transplant, Outcome
Introduction
Tacrolimus is an immunosuppressant isolated from Streptomyces tsukubaensis.1 This drug is effective in preventing organ rejection after kidney, liver, heart, lung, pancreas, small intestine, and bone-marrow transplants.2,3 However, tacrolimus is characterized by considerable individual variability in pharma-cokinetics owing to its irregular intestinal absorption and narrow therapeutic window. Therefore, health care professionals must carefully monitor blood tacrolimus concentrations to avoid an incorrect drug dosage that may cause graft rejection and other adverse effects, such as nephrotoxicity, neurotoxicity, hypertension, and infection—if underdosed or overdosed.4
In our hospital, we measure tacrolimus concentrations using the Dimension Xpand Plus with the Heterogenous Module Integrated Chemistry System (Siemens Corporation, Deerfield, IL, USA). This machine adopts an antibody-conjugated magnetic immunoassay (ACMIA). Antibody-conjugated magnetic immunoassay is a method that measures drug concentrations, including tacrolimus, cyclosporine, and digoxin using a mouse capture antibody, magnetic particles, and β-galactosidase. Several studies have shown that the results of the ACMIA correlate with those of alternative immunoassays.5,6 Additionally, compared with other immunoassays, the ACMIA does not interfere with hematocrit values, and pretreatment is not required. However, in the last decade, some cases have been described in which the blood tacrolimus concentrations as measured by the ACMIA revealed falsely elevated drug concentrations in kidney, liver, and heart transplant recipients.7-10 Fortunately, the values described were remarkably toxic concentrations; therefore, health care professionals immediately noticed the abnormal values were incorrect because of the inconsistency of clinical symptoms. Unlike previous cases, we recently saw cases that showed falsely detected tacrolimus concentrations (although they were within the therapeutic range). Here, we report 2 cases of falsely detected tacrolimus concentrations that did not show abnormally high values within the therapeutic range.
Case 1
In March 2009, a 63-year-old woman underwent a liver transplant because of a
congenital arteriovenous fistula. As treatment for baseline immuno-suppression,
she had taken cyclosporine (50 mg, orally, twice daily) and mycophenolate
mofetil after the transplant for more than 3 years. The whole-blood trough
cyclosporine concentrations were maintained within 80 ng/mL.
In March 2012, the physician routinely requested the pharmacist to measure the trough tacrolimus concentrations, even though she had never taken tacrolimus. The whole-blood tacrolimus concentrations obtained by the ACMIA run on the Dimension Xp and Plus Heterogenous Module system was 2.7 ng/mL (sensitivity of tacrolimus: 1.2 ng/mL; package insert). Because the pharmacist noticed this inconsistency by checking the patient’s prescription history with an electronic medical chart, the pharmacist immediately measured the whole-blood cyclosporine concentration. The whole-blood cyclosporine concentration obtained by the ACMIA was 47.3 ng/mL. Because of these events, we suspected a false detection of tacrolimus/cyclosporine concentrations by the ACMIA as shown in previous reports.7 To check false detection of tacrolimus concentration, we measured tacrolimus concentration in plasma and washed erythrocytes. The value of tacrolimus concentration in plasma and washed erythrocytes was 2.3 ng/mL and zero (Table 1), although approximately 80% of tacrolimus in blood is distributed in erythrocytes.11 In contrast, cyclosporine concentrations in plasma and washed erythrocytes were reasonable (Table 1). Thus, the whole-blood cyclosporine concentration did not reveal a false elevation. Similar results were observed for values collected between May 2012, and July 2012 (Table 1).
Case 2
We evaluated a 75-year-old woman who had undergone a liver transplant in
August 2006 for cirrhosis after chronic infection with hepatitis C virus. After
the transplant, she was treated with tacrolimus 0.2 mg, twice daily (or
sometimes, 0.2 mg once a day) and mycophenolate mofetil. She has been followed
at our hospital and other hospitals once a month, and the tacrolimus
concentration has been monitored in both hospitals. However, the values of the
whole blood tacrolimus concentration as measured by the ACMIA were different in
our hospital (5-10 ng/mL) than those measured by the enzyme-linked immunosorbent
assay (ELISA) in other hospitals (0.5-2.5 ng/mL) (Figure 1). Thus, we suspected
false detection of tacrolimus concentration by the ACMIA. To investigate
method-specific interference, we compared the whole-blood tacrolimus
concentrations obtained by the ACMIA with those obtained by a chemiluminescent
immunoassay (CLIA) using the same sample. Thus, the whole-blood tacrolimus
concentration obtained by the ACMIA and CLIA was 9.2 and 0.8 ng/mL. Next, we
measured tacrolimus concentration in plasma and washed erythrocytes; tacrolimus
concentration in washed erythrocytes was considerably lower than that in plasma
(Table 1). Interestingly, the cyclosporine concentration also was detected in
whole blood and plasma and digoxin in plasma, which also were assayed by the
ACMIA (assay sensitivity of cyclosporine and digoxin was 25 ng/mL and 0.2 ng/mL;
Table 1), even though she has never taken both cyclosporine and digoxin. Similar
to tacrolimus, the cyclosporine concentration in washed erythrocytes was
considerably lower than that in plasma, and the digoxin concentration obtained
by CLIA was within the detection limit (< 0.3 ng/mL).
Discussion
Immunoassays represent a powerful tool for routinely and high-throughput analyses, because quantitative determinations can be obtained with sufficient accuracy and reproducibility. Therefore, many drug concentrations, such as mTOR inhibitors, can be analyzed by the immunoassay as an alternative method to other assays including high-performance liquid chromatography.12 Antibody-conjugated magnetic immunoassay is a newly developed, fully automated assay for monitoring of therapeutic drugs such as tacrolimus, cyclosporine, and digoxin. Compared with other alternative assays such as microparticle enzyme-linked immunoassay, enzyme-multiplied immunoassay, and ELISA, this assay requires only minimal sample preparation, because the instrumentation allows automatic lysis of erythrocytes, bringing several advantages to the routine laboratory—it limits the time required for determination and reduces the need for personnel to perform the assay.
Several medical institutions across the world have adopted the ACMIA for monitoring therapeutic drugs. However, several cases of false abnormally elevated tacrolimus/cyclosporine concentrations measured by the ACMIA in patients who had undergone kidney, liver, or heart transplant have been reported.7-10 A literature review of falsely detected tacrolimus/cyclosporine concentrations as measured by the ACMIA is shown in Table 2. Furthermore, Moscato and associates have reported a systematic analysis of the extent of false-positive results of tacrolimus concentration obtained by the ACMIA that was observed in 10 of 1058 patients (approximately 1%), compared with that obtained by the liquid chromatography-tandem mass spectroscopy reference method.13 These reports implicated the strong possibility that in some patients, the ACMIA provides incorrect tacrolimus concentration values.
With this evidence, a method for avoiding falsely detected tacrolimus/cyclosporine concentrations as measured by the ACMIA should be developed immediately. A recent publication reported an interference with tacrolimus concentrations as measured by the ACMIA because of the presence of the following factors in patient samples: heterophilic antibodies, rheumatoid factor, anti–double-strand DNA antibodies, and the antibody-enzyme (β-galactosidase).7,14-16 Although we did not investigate the reasons in detail, we estimated that the differences between the 2 cases were related to this false elevation. False detection of cyclosporine and digoxin was observed in case 2, but not in case 1. According to the ACMIA measurement principle, it is strongly speculated that false detection of cyclosporine and digoxin in case 2 was because the ACMIA interfered with β-galactosidase, and in case 1, false detection is not caused because an endogenous substitute interacted specifically with tacrolimus. Further study is required to determine the factors involved in this false detection.
Based on the interference factors previously reported,7,14-16 the simplest approach to determine the nature of the observed interference involves measuring tacrolimus concentrations in plasma. In addition, the use of commercially available heterophilic blocking tubes and polyethylene glycol are other options. Furthermore, several reports have proposed estimation of whole-blood tacrolimus concentration by evaluating tacrolimus concentration in washed erythrocytes and hematocrit.7,9 However, calculating the falsely detected tacrolimus concentration is difficult, unless an apparent abnormal range is observed. The approaches mentioned above are available only to check whether tacrolimus concentrations are false-positive or not. Therefore, a useful and effectively modified ACMIA must be developed that avoids not only falsely detected tacrolimus/cyclosporine concentrations, but other drugs as well.
In conclusion, to the best of our knowledge, this is the first report that reveals falsely detected tacrolimus concentrations (albeit within a therapeutic range) as measured by the ACMIA. This is despite the fact that apparent falsely elevated tacrolimus/cyclosporine concentrations have been previously reported. The cases described here should raise the physician’s concern because the ACMIA is associated with the potential hazard of false-positive results of tacrolimus concentrations even though they are within the normal range.
References:

Volume : 12
Issue : 5
Pages : 474 - 478
DOI : 10.6002/ect.2013.0113
From the 1Department of Pharmacy, Kumamoto University Hospital;
and the 2Department of Pediatric Surgery and Transplantation,
Graduate School of Medical Science, Kumamoto University, Kumamoto, Japan
Acknowledgements: The authors declare that they have no conflicts of
interest, and they received no funding for this study.
Corresponding author: Kazuaki Taguchi PhD, Department of Pharmacy,
Kumamoto University Hospital, 1-1-1 Honjo, Chuo-ku, Kumamoto, 860-8556, Japan
Corresponding author: Faculty of Pharmaceutical Sciences, Sojo
University, 4-22-1 Ikeda, Nishi-ku, Kumamoto, 860-0082, Japan
Phone: +81 96 326 3869
Fax: +81 96 326 5048
E-mail:
k-taguchi@ph.sojo-u.ac.jp
Figure 1. Time-Course of Detected Tacrolimus Concentration in Whole Blood, Plasma, and Washed Erythrocytes in Case 1 (A) and Case 2 (B)
Table 1. Tacrolimus, Cyclosporine, and Digoxin Concentrations Obtained Using the ACMIA in the Case 1 and 2 Patients
Table 2. Review of the Literature of Falsely Detected Tacrolimus or Cyclosporine Concentrations Measured by Antibody-Conjugated Magnetic Immunoassay